Published December 2022 | Version v1
Book

Development of CsI based charge particle detector array for nuclear spectroscopy at IUAC

  • 1. Inter University Accelerator Centre, P. O. Box 10502, New Delhi - 110067 (India)

Description

Developmental activities in charge particle detector array (CPDA) for discrete gamma ray spectrometry is presented. The multi-detector array has been designed and developed for the detection of light charge particles such as protons and α-particles generated in a heavy ion induced reaction in coincidence with γ-rays. The array currently consists of 32 CsI(TI) detectors coupled to photo-diode and has been developed to serve as an ancillary detector system for Indian National Gamma Array (INGA). Off-line test of CsI detector was carried out using radioactive sources 137Cs, 60Co, 241Am and 229Th. Energy resolution of 250 keV (for 8.37 MeV α-particles) and 70 keV (1332 keV γ-rays) were observed. The array was tested for light-charged particle – γ-ray coincidence for the system 12C + 28Si at 30 MeV. Detailed test results and activities on instrumentation development and design will be presented

Part of:
Proceedings of the DAE-BRNS symposium on nuclear physics. V. 66

Additional details

Publishing Information

Publisher
Cotton University
Imprint Place
Guwahati (India)
Imprint Title
Proceedings of the DAE-BRNS symposium on nuclear physics. V. 66
Imprint Pagination
[1318 p.]
Journal Page Range
[2 p.]

Conference

Title
66. DAE-BRNS symposium on nuclear physics
Dates
1-5 Dec 2022
Place
Guwahati (India)

Optional Information

Notes
Article No. G13