Published November 1997 | Version v1
Journal article

Using a multiple RF coil to widen FOV and improve sensitivity of head-neck MRI

  • 1. Hitachi Medical Corporation, Chiba (Japan). Research and Development Center

Description

A multiple RF coil consisting of a head QD coil and a neck solenoid coil is developed for a vertical magnetic field MRI system, and its feasibility is demonstrated. The field-of-view (FOV) and sensitivity of the multiple RF coil under object-loaded conditions are calculated. The results show that the multiple RF coil has 1.7 times wider FOV and almost 10% higher sensitivity than a conventional head QD coil. A 0.3-T head-neck multiple RF coil is fabricated and imaging experiments are performed. Combination of the two RF receiving coils' signals leads to a wide FOV for both phantom images and human head-neck images. A dip in the sensitivity profile of the sagittal images relating to the setting position of the RF receiving coil is corrected by using shading correction. Thus, head-neck sagittal images having wider FOV and higher sensitivity are obtained from vertical magnetic field MRI. (author)

Additional details

Publishing Information

Journal Title
Medical Imaging Technology
Journal Volume
15
Journal Issue
6
Journal Page Range
p. 734-741.
ISSN
0288-450X
CODEN
MITEET

INIS

Country of Publication
Japan
Country of Input or Organization
Japan
INIS RN
29007670
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOLOGICAL MODELS; HEAD; IMAGES; NMR IMAGING; PHANTOMS; SENSITIVITY; SIGNALS
Descriptors DEC
BODY; BODY AREAS; DIAGNOSTIC TECHNIQUES; MOCKUP; STRUCTURAL MODELS