Published January 1985
| Version v1
Journal article
Consistent-performance test for quality control in X-ray imaging technology including film processing
Creators
- 1. Bayerisches Landesinstitut fuer Arbeitsschutz, Muenchen (Germany, F.R.)
Description
The consistent-performance test will monitor the standard of technical quality of an X-ray picture facility by means of selected test parameters (dose, optical density, contrast, useful radiation field). Over and beyond, the same test radiograph is used for checking film processing. By dividing the testing equipment into absorption filters for insertion into the collimator and test plate, a high degree of practicability is obtained like e.g. easy mounting on lung wall holders. The test specimen made from PTFE (Teflon) displays largely tissue-equivalent properties when X-rayed and, consequently, may be used as a 'substitute patient' over and beyond consistent-performance testing. (orig.)
Abstract (German)
Der beschriebene Konstanztest ueberwacht den technischen Qualitaetsstandard eines Roentgen-Aufnahmeplatzes anhand einiger ausgewaehlter Pruefgroessen (Dosis, optische Dichte, Kontrast, Nutzstrahlenfeld). Darueberhinaus wird mit der gleichen Testaufnahme die Filmverarbeitung ueberprueft. Durch Zweiteilung der Pruefeinrichtung in Absorptionsfilter zum Einschieben in das Lichtvisier und Testplatte wird ein hohes Mass an Praktikabilitaet erreicht, wie z.B. leichte Anbringung an Lungen-Wandstativen. Der Pruefkoerper aus dem Kunststoff PTFE (Teflon) zeigt gegenueber Roentgenstrahlung weitgehend gewebeaequivalente Eigenschaften und kann daher ueber die Konstanzpruefung hinaus als 'Ersatzpatient' verwendet werden. (orig.)Additional details
Additional titles
- Original title (German)
- Ein Konstanztest zur Qualitaetssicherung in der Roentgen-Aufnahmetechnik einschliesslich der Filmverarbeitung
Publishing Information
- Journal Title
- Roentgenpraxis
- Journal Volume
- 38
- Journal Issue
- 1
- Series
- CODEN: RGPXB.;Roentgenpraxis.
- Journal Page Range
- 22-29
- ISSN
- 0035-7820
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 16045619
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DENSITY MATRIX; PHANTOMS; QUALITY CONTROL; RADIATION DOSES; RADIATION MONITORING; RADIATION PROTECTION; RADIATION PROTECTION LAW; RECOMMENDATIONS; X-RAY EQUIPMENT
- Descriptors DEC
- CONTROL; DIAGNOSTIC TECHNIQUES; EQUIPMENT; LAW; MATRICES; MEDICINE; MOCKUP; MONITORING; STRUCTURAL MODELS