Published January 1982 | Version v1
Journal article

Mould therapy using cathetron source pencils

Creators

  • 1. Bhabha Atomic Research Centre, Bombay (India). Div. of Radiological Protection

Description

The use of Cathetron source pencils for planar, linear and cylindrical mould applications has been worked out. Since the source distributions based on Paterson-Parker rules involve the use of the same source pencil for a number of times at different positions on the mould, an alternate arrangement has been worked out which makes use of more than one type of pencil, thereby minimizing the irradiation time. Dose distributions have been obtained for three sizes of planar moulds and two sizes each of linear and cylindrical moulds for different treating distances. (orig.)

Abstract (German)

Die Anwendung von Cathetron-Quellenstiften wurde fuer die Benutzung linearer, zylindrischer und flaechiger Moulagen berechnet. Da die Quellenverteilungen auf der Grundlage der Paterson-Parker-Regeln den mehrmaligen Gebrauch des gleichen Quellenstifts in verschiedenen Stellungen auf der Moulage voraussetzen, diente eine alternierende Anordnung, bei der mehr als ein Typus Quellenstift benutzt wird, zur Berechnung, wodurch sich die Bestrahlungszeit erheblich verkuerzt. Die Dosisverteilungen wurden auf diese Art fuer drei Plattenmoulagen verschiedener Groesse und fuer je zwei Groessen linearer und zylindrischer Moulagen bei unterschiedlichen Behandlungsabstaenden festgestellt. (orig.)

Additional details

Publishing Information

Journal Title
Strahlentherapie
Journal Volume
158
Journal Issue
1
Series
Strahlentherapie.
Journal Page Range
34-39
ISSN
0039-2073