Published 2006 | Version v1
Miscellaneous Restricted

Refinement of the dual ionisation chamber dosimetry carried out at the accelerator-based epithermal neutron beam facility of the University of Birmingham

  • 1. Dept. of Medical Physics, QE University Hospital, Birmingham (United Kingdom)
  • 2. School of Physics and Astronomy, Univ. of Birmingham, Birmingham (United Kingdom)

Description

The paper presents the refined dual ionisation chamber technique used for in-air and in-phantom measurements in the Birmingham epithermal neutron beam. The study includes the derivation of the spectrum-dependent relative neutron sensitivity of the tissue-equivalent ionisation chamber. The average values over shallow depths for the kt parameter in A150 is 0.85 +/- 0.04, corresponding to an average value of 0.80 for water. For photon dosimetry in mixed fields, the formalism initially proposed by Munck af Rosenschold et al has been applied at a specific depth of 3 cm using MCNP4C as the radiation transport tool in the mixed beam and the reference calibration beam to generate electron fluence profiles in the detector gas cavities. The BEAMnrc code was used to generate the starting photon spectrum for the 8MV photon beam. The effect of the chosen energy-indexing algorithm on the in-cavity electron dose using the MNCP4C *F8 tally was also investigated. (author)

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Part of:
Advances in neutron capture therapy 2006. Proceedings of 12th international congress on neutron capture therapy

Additional details

Publishing Information

Imprint Title
Advances in neutron capture therapy 2006. Proceedings of 12th international congress on neutron capture therapy
Imprint Pagination
638 p.
Journal Page Range
p. 445-448
Report number
INIS-JP--103

Conference

Title
12. international congress on neutron capture therapy
Acronym
ICNCT-12
Dates
9-13 Oct 2006
Place
Takamatsu, Kagawa (Japan)

Optional Information

Notes
9 refs., 1 fig., 2 tabs.