Published June 1994
| Version v1
Journal article
Standardization and half-life of 201Tl by the 4π(x,e)-γ coincidence method with liquid scintillation counting in the 4π-channel
Description
A high-quality solution of 201Tl has been standardized by the 4π(x,e)-γ coincidence method in conjunction with efficiency extrapolation. Liquid scintillation was utilized for the detection of the X-rays and electrons in the 4π-channel. The accuracy of the technique was verified by submitting a standardized sample to the International Reference System (SIR) for comparison with activity results obtained from other methods. The decay of one of the sources was followed over a 2-week period by measuring the activity with this direct method on 91 occasions. The resulting 201Tl half-life was (3.0400 ± 0.0028) days. (Author)
Additional details
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 45
- Journal Issue
- 6
- Journal Page Range
- p. 669-673.
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 25052296
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- ACCURACY; CALIBRATION STANDARDS; COINCIDENCE METHODS; FOUR-PI COUNTING; HALF-LIFE; LIQUID SCINTILLATION DETECTORS; THALLIUM 201
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; COUNTING TECHNIQUES; DAYS LIVING RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; HEAVY NUCLEI; ISOTOPES; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; SCINTILLATION COUNTERS; STANDARDS; THALLIUM ISOTOPES