Published November 1984
| Version v1
Journal article
Computer-aided design of a new high-performance mass spectrometer
Description
A new procedure for designing a high-performance mass spectrometer is described. TRIO-FIT, a computer program which utilizes an automatic fitting function, is employed for the purpose. As an example of TRIO-FIT's usefulness, an ion-optical design for a high-performance mass spectrometer (high resolution, mass range, and transmission) is developed from several alternatives. A mass resolution of 300,000 and a mass range of up to M/Z 200,000 can be expected from this system. A method of examining the practicability of a given mass spectrometer design is also described, and this procedure is demonstrated on the proposed design of the high performance instrument. (author)
Additional details
Publishing Information
- Journal Title
- Jpn. J. Appl. Phys., Part 1
- Journal Volume
- 23
- Journal Issue
- 11
- Series
- Jpn. J. Appl. Phys., Part 1.
- Journal Page Range
- 1518-1525
- CODEN
- JAPND
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 17007209
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BEAM OPTICS; COMPUTER-AIDED DESIGN; DATA PROCESSING; ELECTRIC FIELDS; IONS; MAGNETIC FIELDS; MASS RESOLUTION; MASS SPECTROMETERS; T CODES; TRANSFER MATRIX METHOD; TRANSMISSION
- Descriptors DEC
- CHARGED PARTICLES; COMPUTER CODES; MEASURING INSTRUMENTS; RESOLUTION; SPECTROMETERS