Consistency check of photon beam physical data after recommissioning process
- 1. Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai (Thailand)
- 2. Lampang Cancer Hospital, Lampang (Thailand)
Description
In radiotherapy, medical linear accelerator (Linac) is the key system used for radiation treatments delivery. Although, recommissioning was recommended after major modification of the machine by AAPM TG53, but it might not be practical in radiotherapy center with heavy workloads. The main purpose of this study was to compare photon beam physical data between initial commissioning and recommissioning of 6 MV Elekta Precise linac. The parameters for comparing were the percentage depth dose (PDD) and beam profiles. The clinical commissioning test cases followed IAEA-TECDOC-1583 were planned on REF 91230 IMRT Dose Verification Phantom by Philips' Pinnacle treatment planning system. The Delta4PT was used for dose distribution verification with 90% passing criteria of the gamma index (3%/3mm). Our results revealed that the PDDs and beam profiles agreed within a tolerance limit recommended by TRS430. Most of the point doses and dose distribution verification passed the acceptance criteria. This study showed the consistency of photon beam physical data after recommissioning process. There was a good agreement between initial commissioning and recommissioning within a tolerance limit, demonstrated that the full recommissioning process might not be required. However, in the complex treatment planning geometry, the initial data should be applied with great caution. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/694/1/012023Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 694
- Journal Issue
- 1
- Journal Page Range
- [5 p.]
- ISSN
- 1742-6596
Conference
- Title
- 13. South-East Asian congress of medical physics (SEACOMP) 2015
- Dates
- 10-12 Dec 2015
- Place
- Yogyakarta (Indonesia)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47117950
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BEAM PROFILES; COMPARATIVE EVALUATIONS; DEPTH; DEPTH DOSE DISTRIBUTIONS; IAEA; LINEAR ACCELERATORS; MODIFICATIONS; PHANTOMS; PHOTON BEAMS; PHOTONS; PLANNING; RADIATION DOSES; RADIOTHERAPY; TOLERANCE; VERIFICATION
- Descriptors DEC
- ACCELERATORS; BEAMS; BOSONS; DIMENSIONS; DOSES; ELEMENTARY PARTICLES; EVALUATION; INTERNATIONAL ORGANIZATIONS; MASSLESS PARTICLES; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS; THERAPY