Published November 1992 | Version v1
Journal article

The use of Tri-Carb 1050 TR/LL liquid scintillation counter for the measurement of tritium contamination in environmental waters

  • 1. Ceskoslovenska Akademie Ved, Prague (Czech Republic). Ustav Radiologicke Dozimetrie

Description

The use of the Tri-Carb 1050 TR/LL liquid scintillation spectrometer for the analysis of low tritium concentration in environmental samples is described. Measurement optimization is discussed and the sensitivity and accuracy of counting are demonstrated on the determination of HTO content of various natural water samples. The minimum detectable activity of (1.2 Bq.l-1)0.95 and 1000 minutes' counting time guarantee a reliable assessment of low tritium contamination from the nuclear industry and make this instrument suitable for monitoring and control of water supplies pollution. (author) 1 tab., 4 figs., 10 refs

Additional details

Publishing Information

Journal Title
Jaderna Energie
Journal Volume
38
Journal Issue
11
Journal Page Range
p. 405-408.
ISSN
0448-116X
CODEN
JADEAQ