Published January 1997
| Version v1
Journal article
A note on fabrication and performance test of a proportional counter for soft X-ray spectroscopy
- 1. Tokai Univ., Hiratsuka, Kanagawa (Japan). Faculty of Engineering
Description
A proportional counter has been designed to detect soft X-rays in energy-dispersive X-ray fluorescence (EDXRF) analysis of elements having a medium atomic number. The counter consists of a 25 mm diameter cylindrical cathode with an entrance window of a thin Mylar film and a 50 μm diameter anode wire, it was filled with PR gas at 1 atm using by a gas flow system. The maximum gas multiplication of the counter is over 4 x 104 at 2000 V of the applied voltage and the energy resolution for X-ray is between 17% and 35% in the energy range between 5.9 keV and 15 keV. The counter exceeds PIN photo-diode Si detector (AMPTEK's type XR-100) in detection efficiency under 15 keV of the X-ray energy. (author)
Additional details
Publishing Information
- Journal Title
- Radioisotopes (Tokyo)
- Journal Volume
- 46
- Journal Issue
- 1
- Journal Page Range
- p. 28-32.
- ISSN
- 0033-8303
- CODEN
- RAISAB
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 28033005
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- ENERGY RESOLUTION; GAS FLOW; IRON 55; PROPORTIONAL COUNTERS; X RADIATION; X-RAY DETECTION; X-RAY FLUORESCENCE ANALYSIS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; CHEMICAL ANALYSIS; DETECTION; ELECTROMAGNETIC RADIATION; ELECTRON CAPTURE RADIOISOTOPES; EVEN-ODD NUCLEI; FLUID FLOW; INTERMEDIATE MASS NUCLEI; IONIZING RADIATIONS; IRON ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; NONDESTRUCTIVE ANALYSIS; NUCLEI; RADIATION DETECTION; RADIATION DETECTORS; RADIATIONS; RADIOISOTOPES; RESOLUTION; X-RAY EMISSION ANALYSIS; YEARS LIVING RADIOISOTOPES