Published January 1997 | Version v1
Journal article

A note on fabrication and performance test of a proportional counter for soft X-ray spectroscopy

  • 1. Tokai Univ., Hiratsuka, Kanagawa (Japan). Faculty of Engineering

Description

A proportional counter has been designed to detect soft X-rays in energy-dispersive X-ray fluorescence (EDXRF) analysis of elements having a medium atomic number. The counter consists of a 25 mm diameter cylindrical cathode with an entrance window of a thin Mylar film and a 50 μm diameter anode wire, it was filled with PR gas at 1 atm using by a gas flow system. The maximum gas multiplication of the counter is over 4 x 104 at 2000 V of the applied voltage and the energy resolution for X-ray is between 17% and 35% in the energy range between 5.9 keV and 15 keV. The counter exceeds PIN photo-diode Si detector (AMPTEK's type XR-100) in detection efficiency under 15 keV of the X-ray energy. (author)

Additional details

Publishing Information

Journal Title
Radioisotopes (Tokyo)
Journal Volume
46
Journal Issue
1
Journal Page Range
p. 28-32.
ISSN
0033-8303
CODEN
RAISAB