Published November 1986
| Version v1
Journal article
Line spread functions by the pre-exposure and multiple-exposure methods
Creators
- 1. Tokoha Gakuen Univ., Sena, Shizuoka (Japan). Science Dept.
- 2. Miyazaki Medical Coll., Kiyotake (Japan)
- 3. Gifu Coll. of Medical Technology (Japan)
- 4. Gifu Univ. (Japan). Faculty of Engineering
Description
The multiple-exposure method is widely used to measure line spread functions (LSF) of screen-film combinations. However, it is not easy for everyone to use this method, because it is necessary to combine many curves obtained by individual experiments graphically. A new and useful method to measure LSF by making use of pre-exposure techniques is presented in this paper. Compared with the multiple-exposure method, the new method is simple and useful. It is shown that LSF obtained by the new method agree very well with those obtained by the multiple-exposure method. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 31
- Journal Issue
- 11
- Series
- Phys. Med. Biol.
- Journal Page Range
- 1275-1279
- ISSN
- 0031-9155
- CODEN
- PHMBA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 18023083
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; IMAGE INTENSIFIERS; IMAGE PROCESSING; PHOTOGRAPHIC FILMS; RESOLUTION; X RADIATION
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; RADIATIONS