Published November 1987
| Version v1
Journal article
Neutron REM counter
- 1. Fuji Electric Co. Ltd., Tokyo (Japan)
Description
High sensitivity, light weight 'neutron rem counter' was developed by using 3He proportional counter. It was confirmed that the response to neutron energy from thermal to 8 MeV is well fitted to the ICRP-21 rem response. The neutron sensitivity of this is 50 cps/mremh-1 and the weight is about 6.5 kg. This report also gives on outline of neutron dose measurement. (author)
Additional details
Publishing Information
- Journal Title
- FAPIG (Tokyo)
- Journal Issue
- no.117
- Series
- FAPIG (Tokyo).
- ISSN
- 0014-5645
- CODEN
- FPIGA
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 20003956
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COUNTING TECHNIQUES; DOSE EQUIVALENTS; ELECTRONIC CIRCUITS; ENERGY DEPENDENCE; MODERATING DETECTORS; NEUTRON MONITORS; PERFORMANCE; RESPONSE FUNCTIONS; SPECIFICATIONS
- Descriptors DEC
- FUNCTIONS; MEASURING INSTRUMENTS; MONITORS; NEUTRON DETECTORS; RADIATION DETECTORS; RADIATION MONITORS