Published February 2015 | Version v1
Journal article

On-site calibration technology for fixed environmental gamma radiation ratemeter

  • 1. China Institute of Atomic Energy, Beijing (China)

Description

The fixed environmental gamma radiation ratemeter is the main equipment to undertake environmental continuous monitoring tasks and not easy to move for metrology laboratory calibration, and the submission of a longer period affects the continuity of the continuous monitoring data. For the on time calibration of fixed environmental gamma radiation ratemeter, combined with the Monte Carlo method, the energy-compensated high-pressure ionization chamber and portable 137Cs irradiation facility were developed. Using natural background radiation (including cosmic rays and terrestrial gamma rays) and 137Cs gamma rays reference radiation field, the on-site calibration for fixed environmental gamma radiation ratemeter was carried out. The results show that the environment intercomparison and on-site irradiation can solve the calibration problem of fixed environmental gamma radiation ratemeters, and the relative deviation of calibration factors obtained from on-site calibration and in standard laboratory is less than 5%. (authors)

Additional details

Identifiers

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
49
Journal Issue
2
Journal Page Range
p. 212-217
ISSN
1000-6931

Conference

Title
28. May Four Young Scientists Symposium Held in China Institute of Atomic Energy
Dates
2014
Place
Beijing (China)

Optional Information

Notes
4 figs., 4 tabs., 12 refs.; http://dx.doi.org/10.7538/yzk.2015.49.02.0212