Published June 2008 | Version v1
Journal article

Measurement of the D-D neutron generation rate by proton counting

  • 1. Seoul National University, Seoul (Korea, Republic of)

Description

A detection system was set up to measure the neutron generation rate of a recently developed D-D neutron generator. The system is composed of a Si detector, He-3 detector, and electronics for pulse height analysis. The neutron generation rate was measured by counting protons using the Si detector, and the data was crosschecked by counting neutrons with the He-3 detector. The efficiencies of the Si and He-3 detectors were calibrated independently by using a standard alpha particle source 241Am and a bare isotopic neutron source 252Cf, respectively. The effect of the cross-sectional difference between the D(d,p)T and D(d,n)3He reactions was evaluated for the case of a thick target. The neutron generation rate was theoretically corrected for the anisotropic emission of protons and neutrons in the D-D reactions. The attenuations of neutron on the path to the He-3 detector by the target assembly and vacuum flange of the neutron generator were considered by the Monte Carlo method using the MCNP 4C2 code. As a result, the neutron generation rate based on the Si detector measurement was determined with a relative uncertainty of ±5%, and the two rates measured by both detectors corroborated within 20%

Additional details

Publishing Information

Journal Title
Nuclear Engineering and Technology
Journal Volume
40
Journal Issue
4
Series
16 refs, 9 figs, 1 tab
Journal Page Range
p. 299-304
ISSN
1738-5733