Published August 1981 | Version v1
Journal article

Determination of small amounts of aluminum in iron and steel by x-ray fluorescence analysis

  • 1. Daido Steel Co. Ltd., Nagoya (Japan). Central Research Lab.

Description

The accuracy of X-ray fluorescence analysis for determining aluminum in steel is generally poor when the aluminum content is below 0.1%. In this study, it is confirmed that the existence of inhomogeneous acid-insoluble aluminum causes the poor accuracy of X-ray fluorescence analysis. The accuracy can be improved by adding granular ferro-zirconium to molten steel when sampling. By using this method, the accuracy of determining aluminum was 0.0024% for low alloy steels and 0.0022% for stainless steels as the standard deviation obtained for a series of differences between an analytical result and a standard value. Corrections for overlapping of the chromium spectrum, and for absorption and enhancement effects of co-existent elements are not necessary for the analysis of low alloy steels, whereas these corrections are necessary for the analysis of stainless steels. (author)

Additional details

Publishing Information

Journal Title
Tetsu To Hagane
Journal Volume
67
Journal Issue
10
Series
Tetsu To Hagane.
Journal Page Range
1823-1828