Precision of the gallbladder ejection fraction obtained with Tc-99m-pyridoxyl-5-methyl-tryptophan (99mTc-PMT) hepatobiliary scintigraphy as compared with the contraction ratio in three-dimensional computed tomography

  • 1. Showa University, Yokohama (Japan). Fujigaoka Hospital

Description

The gallbladder ejection fraction (GBEF) obtained with Tc-99m-pyridoxyl-5-methyl-tryptophan (99mTc-PMT) hepatobiliary scintigraphy has been used as a parameter of gallbladder function. To determine the accuracy of GBEF, the relationship with the contraction ratio of the gallbladder (GBCR) obtained with three-dimensional helical computed tomography (3D-CT) was studied. A normal volunteer, 8 patients suffering from cholecystolithiasis and a patient with gallbladder dyskinesia were examined. The percent initial dose (%ID) for the gallbladder and GBEF with hepatobiliary scintigraphy were used to compare the volume of the gallbladder and GBCR which was measured by 3D-CT. The %ID of the gallbladder was correlated with the volume of the gallbladder by 3D-CT (Y=1.000X-1.818, r=0.928). GBEF was correlated well with GBCR by 3D-CT (Y=0.916X + 6.296, r=0.975). The %ID of the gallbladder obtained with hepatobiliary scintigraphy may be a good indicator of the volume of the gallbladder. The accuracy of GBEF was confirmed by comparison with 3D-CT examination. GBEF is considered a useful parameter of pathophysiological gallbladder function. (author)