Published April 2007 | Version v1
Journal article

Use of a SQUID array to detect T-cells with magnetic nanoparticles in determining transplant rejection

  • 1. Senior Scientific, 11109 Country Club NE, Albuquerque, NM 87111 (United States)
  • 2. Department of Physics and Astronomy, University of New Mexico, Albuquerque, NM 87131 (United States)
  • 3. Chemicell GmbH, Berlin (Germany)
  • 4. Cancer Research and Treatment Center, University of New Mexico, Albuquerque, NM 87131 (United States)

Description

Acute rejection in organ transplant is signaled by the proliferation of T-cells that target and kill the donor cells requiring painful biopsies to detect rejection onset. An alternative non-invasive technique is proposed using a multi-channel superconducting quantum interference device (SQUID) magnetometer to detect T-cell lymphocytes in the transplanted organ labeled with magnetic nanoparticles conjugated to antibodies specifically attached to lymphocytic ligand receptors. After a magnetic field pulse, the T-cells produce a decaying magnetic signal with a characteristic time of the order of a second. The extreme sensitivity of this technique, 105 cells, can provide early warning of impending transplant rejection and monitor immune-suppressive chemotherapy

Additional details

Identifiers

DOI
10.1016/j.jmmm.2006.10.1148;
PII
S0304-8853(06)02532-7;

Publishing Information

Journal Title
Journal of Magnetism and Magnetic Materials
Journal Volume
311
Journal Issue
1
Journal Page Range
p. 429-435
ISSN
0304-8853
CODEN
JMMMDC

Conference

Title
6. international conference on the scientific and clinical applications of magnetic carriers
Acronym
SCAMC-06
Dates
17-20 May 2006
Place
Krems (Austria)

Optional Information