Gamma sensitivity of the Eberline PCM-1
Description
This paper reports that normally, alarm setpoints for the Eberline PCM-1 series personnel contamination monitors are calculated based upon efficiencies measured using 100-cm2 or larger beta or mixed beta-gamma sources. This simulates the type of contamination most frequently encountered on personnel and clothing--low-level, distributed beta-gamma emitters. In most circumstances, the PCM-1's sensitivity to the other type of contamination encountered in nuclear plant work--hot particles--would be expected to be the same as, or better than, its sensitivity to distributed contamination. However, particles that are deposited on skin or clothing can be partially shielded from the view of the PCM-1's detectors. In these situations, the PCm-1's sensitivity to gamma radiation may be more relevant than its sensitivity to betas
Additional details
Publishing Information
- Journal Title
- Radiation Protection Management
- Journal Volume
- 5
- Journal Issue
- 6
- Series
- Radiat. Prot. Manage.
- Journal Page Range
- 43-46
- ISSN
- 0740-0640
- CODEN
- RPMAE
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 23071504
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BETA SOURCES; CALCULATION METHODS; CLOTHING; CONTAMINATION; DOSIMETRY; GAMMA SOURCES; MIXTURES; PERSONNEL; RADIATION DOSES; RADIATION MONITORING; RADIATION MONITORS; SENSITIVITY; SKIN
- Descriptors DEC
- BODY; DISPERSIONS; MEASURING INSTRUMENTS; MONITORING; MONITORS; ORGANS; PARTICLE SOURCES; RADIATION SOURCES