AMS of heavy elements with an ECR ion source and the ATLAS linear accelerator

Description

Detection of heavy elements by accelerator mass spectrometry with the electron cyclotron resonance ion source, Argonne linear accelerator and fragment mass analyzer (ECRIS-ATLAS-FMA) system has been developed. The use of the ECR-ATLAS system for AMS of heavy elements has two interesting features: (i) the efficient production of high-charge state ions in the ECR source ensures the elimination of molecular ions at the source stage, a highly attractive feature for any mass-spectrometric use not exploited so far; (ii) the linear acceleration based on velocity matching and the beam transport system act as a powerful mass filter for background suppression. We have shown that our system reaches an abundance sensitivity of 1x10-14 for Pb isotopes. The 236U detection sensitivity is 236U/U > or approx. 1x10-12, limited mainly by the ion source output

Additional details

Identifiers

PII
S0168583X00002792;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
Journal Volume
172
Journal Issue
1-4
Journal Page Range
p. 688-692
ISSN
0168-583X
CODEN
NIMBEU