Published April 2009 | Version v1
Journal article

Air-kerma cavity standards

  • 1. Physikalisch-Technische Bundesanstalt, Braunschweig (Germany)
  • 2. Bureau International des Poids et Mesures, 92 - Sevres (France)

Description

This paper reviews the current status of air-kerma cavity standards for gamma rays from 60Co, 137Cs and 192Ir sources. It describes the basic principles and the theories to determine the air-kerma rate from the ionization current of graphite-walled cavity ionization chambers. A typical design of a cavity chamber is shown and described. The various correction factors to be applied to the standards and their determination by experimental and Monte Carlo methods are discussed. A typical uncertainty budget for a cavity standard is presented and the results of comparisons between primary standards are summarized and discussed. Finally, conclusions are drawn and proposals made for topics to be addressed in future studies. (authors)

Availability note (English)

Available from doi:

Additional details

Identifiers

Publishing Information

Journal Title
Metrologia
Journal Volume
46
Journal Issue
no.2
Journal Page Range
p. S24-S38
ISSN
0026-1394
CODEN
MTRGAU

Optional Information

Notes
90 refs.