Expected resolution of plutonium-241 in electron microscope autoradiography
Creators
Description
The error in location of an autoradiographic image is affected by the size of the developed grains (photographic error), the thickness of the section and emulsion, and the energy of the radioisotope used (geometric error). The expected resolution of 241Pu is estimated from theoretical considerations and experimentally derived data from other radioisotopes. Using standard autoradiographic procedures, the half-distance value of 241Pu would be about 150 nm and the total autoradiographic error about 180 nm. An error of this magnitude should permit organellar localization of 241Pu in electron microscope autoradiography
Additional details
Publishing Information
- Imprint Title
- Research in radiobiology. Annual report of work in progress in the Internal Irradiation Program, 31 March 1981 and 1982
- Journal Page Range
- p. 169-176.
- Report number
- DOE/EV/00119--257
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14792176
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AUTORADIOGRAPHY; BEAGLES; ELECTRON MICROSCOPY; FORECASTING; GUINEA PIGS; PLUTONIUM 241; RATS; SPATIAL RESOLUTION; STATISTICS; TISSUES
- Descriptors DEC
- ACTINIDE NUCLEI; ALPHA DECAY RADIOISOTOPES; ANIMALS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; DOGS; EVEN-ODD NUCLEI; HEAVY NUCLEI; ISOTOPES; MAMMALS; MATHEMATICS; MICROSCOPY; NUCLEI; PLUTONIUM ISOTOPES; RADIOISOTOPES; RESOLUTION; RODENTS; VERTEBRATES; YEARS LIVING RADIOISOTOPES