EIT Imaging of Upper Airway to Estimate Its Size and Shape Changes During Obstructive Sleep Apnea
- 1. Kyung Hee University, Department of Biomedical Engineering, Graduate School (Korea, Republic of)
- 2. Kyung Hee University, Department of Medical Engineering, Graduate School (Korea, Republic of)
- 3. Gyeongsang National University College of Medicine and Gyeongsang National University Hospital, Department of Otorhinolaryngology (Korea, Republic of)
Description
Noninvasive continuous imaging of the upper airway during natural sleep was conducted for patients with obstructive sleep apnea (OSA) using the electrical impedance tomography (EIT) technique. A safe amount of alternating current (AC) was injected into the lower head through multiple surface electrodes. Since the air is an electrical insulator, upper airway narrowing during OSA altered internal current pathways and changed the induced voltage distribution. Since the measured voltage data from the surface of the lower head were influenced not only by upper airway narrowing but respiratory motions, head motions, and blood flows, we developed a pre-processing algorithm to extract the voltage component originated from upper airway closing and opening. Using an EIT image reconstruction algorithm, time-series of EIT images of the upper airway were produced with a temporal resolution of 50 frames per second. Applying a postprocessing algorithm to the reconstructed EIT images, we could extract quantitative information about changes in the size and shape during upper airway closing and opening. Results of the clinical studies with seven normal subjects and ten OSA patients show the feasibility of the new method for OSA phenotyping and treatment planning.
Additional details
Identifiers
Publishing Information
- Journal Title
- Annals of Biomedical Engineering
- Journal Volume
- 47
- Journal Issue
- 4
- Journal Page Range
- p. 990-999
- ISSN
- 0090-6964
- CODEN
- ABMECF
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54061740
- Subject category
- S60: APPLIED LIFE SCIENCES; S47: OTHER INSTRUMENTATION;
- Descriptors DEI
- ALGORITHMS; ALTERNATING CURRENT; BLOOD FLOW; ELECTRIC POTENTIAL; ELECTRICAL INSULATORS; ELECTRODES; HEAD; IMAGE PROCESSING; IMAGES; IMPEDANCE; PATIENTS; PHENOTYPE; TOMOGRAPHY
- Descriptors DEC
- BODY; CURRENTS; DIAGNOSTIC TECHNIQUES; ELECTRIC CURRENTS; ELECTRICAL EQUIPMENT; EQUIPMENT; MATHEMATICAL LOGIC; PROCESSING
Optional Information
- Copyright
- Copyright (c) 2019 Biomedical Engineering Society