Ultrasound Imaging. Chapter 13

  • 1. University of Western Ontario, London (Canada)

Description

In the conventional method of ultrasonography, images are acquired in reflection, or pulse echo, mode. An array of small piezoelectric elements transmits a focused pulse along a specified line of sight known as a scan line. Echoes returning from the tissue are received by the same array, focused via the delay-and-sum beam forming process reviewed in Section 13.2, and demodulated to obtain the magnitude, or envelope, of the echo signal. The scanner measures the arrival time of the echoes relative to the time the pulse was transmitted and maps the arrival time to the distance from the array, using an assumed speed of sound. The earliest ultrasound systems would display the result of a single pulse acquisition in 1-D A-mode (amplitude mode) format by plotting echo magnitude as a function of distance. A 2-D or 3-D B-mode (brightness mode) image is acquired by performing a large number of pulse echo acquisitions, incrementally increasing the scan line direction between each pulse echo operation, to sweep out a 2-D or 3-D field of view (FOV). The term B-mode imaging reflects the fact that the echo magnitude from each point in the FOV is mapped to the grey level, or brightness, of the corresponding pixel in the image

Part of:
Diagnostic radiology physics: A handbook for teachers and students. Endorsed by: American Association of Physicists in Medicine, Asia-Oceania Federation of Organizations for Medical Physics, European Federation of Organisations for Medical Physics

Additional details

Publishing Information

Publisher
IAEA
Imprint Place
Vienna (International Atomic Energy Agency (IAEA))
ISBN
978-92-0-131010-1
Imprint Title
Diagnostic radiology physics: A handbook for teachers and students. Endorsed by: American Association of Physicists in Medicine, Asia-Oceania Federation of Organizations for Medical Physics, European Federation of Organisations for Medical Physics
Imprint Pagination
710 p.
Journal Page Range
p. 311-331

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46065930
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
AMPLITUDES; ANIMAL TISSUES; FORMATES; IMAGE PROCESSING; IMAGES; PIEZOELECTRICITY; REVIEWS; ULTRASONOGRAPHY
Descriptors DEC
BODY; CARBOXYLIC ACID SALTS; DIAGNOSTIC TECHNIQUES; DOCUMENT TYPES; ELECTRICITY; PROCESSING

Optional Information

Notes
Figs.
Secondary number(s)
STI/PUB--1564