Published July 21, 2001
| Version v1
Journal article
Recent performance and developments on the Daresbury SRS soft X-ray undulator beamline
Creators
Description
The recent performance of the Daresbury Laboratory soft X-ray undulator beamline 5U1 is presented. The beamline resolution has been estimated from the photo-absorption of various gases. The resolution does not follow the expected trend with photon energy due to slope error problems with the optics. The flux has been measured by the drain-current from a gold photodiode and is in good agreement with the theoretical prediction. The performance of a new ellipsoidal post-focusing mirror is found to be in excellent agreement with the predictions from the ray tracing. The effect of movement of the source on photon energy calibration is considered
Additional details
Identifiers
- PII
- S0168900201003989;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 467-468
- Journal Issue
- 1-2
- Journal Page Range
- p. 516-519
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Germany
- INIS RN
- 33062799
- Subject category
- S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- BEAM TRANSPORT; CALIBRATION; ENERGY RESOLUTION; MIRRORS; ONDULATOR RADIATION; PHOTON BEAMS; RADIATION FLUX; SOFT X RADIATION; SYNCHROTRON RADIATION SOURCES
- Descriptors DEC
- BEAMS; BREMSSTRAHLUNG; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; RADIATION SOURCES; RADIATIONS; RESOLUTION; X RADIATION
Optional Information
- Copyright
- Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.