Published July 15, 1992
| Version v1
Journal article
Single- to double-escape peak ratio of germanium detectors
Creators
- 1. Nuclear Science and Technology Development Center, National Tsing Hua Univ., Hsinchu (Taiwan)
Description
Twenty-one HPGe detectors with a crystal volume of 60-256 cm3 were used to explore the dependence of the single- to double-escape peak ratio on the detector size. The variations of the escape peak ratio as functions of both relative efficiency and crystal volume of HPGe detectors are quantitatively described and the geometric influence on the ratio is also discussed for the coaxial detectors with the D/H ratio within the range of 0.85-1.44. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A
- Journal Volume
- 317
- Journal Issue
- 3
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 537-540
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 23087532
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- EFFICIENCY; ESCAPE PEAKS; GAMMA DETECTION; GEOMETRY; HIGH-PURITY GE DETECTORS; MEV RANGE 01-10
- Descriptors DEC
- DETECTION; ENERGY RANGE; GE SEMICONDUCTOR DETECTORS; MATHEMATICS; MEASURING INSTRUMENTS; MEV RANGE; PEAKS; RADIATION DETECTION; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS