Published July 2008 | Version v1
Journal article

Design and construction of a Cerenkov counter for in situ monitoring of 90Sr in groundwater

  • 1. Pacific Northwest National Laboratory, Richland, WA (United States)

Description

Migration of groundwater contamination from beneath the Hanford Site into the Columbia River creates a need for in situ 90Sr monitoring. The prototype monitor discussed here was designed for deployment downwell and provides near-real-time determination of the 90Sr concentration in a twoliter groundwater sample. This manuscript discusses the design, construction and preliminary results for a prototype monitor. Calibration and testing results of the as-built system show qualitative agreement between simulated predictions and experimental results. Downwell and laboratory tests demonstrate a sensitivity to 90Sr at concentrations below 8 pCi/l (0.3 Bq/l) at a 90% confidence level in measurement times of less than four hours. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
277
Journal Issue
1
Journal Page Range
p. 235-239
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
4 refs.