Performance measurements of a depth-encoding PET detector module based on position-sensitive avalanche photodiode read-out
Creators
- 1. Department of Biomedical Engineering, University of California-Davis, One Shields Avenue, Davis, CA 95616 (United States)
- 2. Department of Molecular and Medical Pharmacology, UCLA, Los Angeles, CA 90095 (United States)
- 3. Radiation Monitoring Devices Inc., Watertown, MA 02172 (United States)
Description
We are developing a high-resolution, high-efficiency positron emission tomography (PET) detector module with depth of interaction (DOI) capability based on a lutetium oxyorthosilicate (LSO) scintillator array coupled at both ends to position-sensitive avalanche photodiodes (PSAPDs). In this paper we present the DOI resolution, energy resolution and timing resolution results for complete detector modules. The detector module consists of a 7 x 7 matrix of LSO scintillator crystals (1 x 1 x 20 mm3 in dimension) coupled to 8 x 8 mm2 PSAPDs at both ends. Flood histograms were acquired and used to generate crystal look-up tables. The DOI resolution was measured for individual crystals within the array by using the ratio of the signal amplitudes from the two PSAPDs on an event-by-event basis. A measure of the total scintillation light produced was obtained by summing the signal amplitudes from the two PSAPDs. This summed signal was used to measure the energy resolution. The DOI resolution was measured to be 3-4 mm FWHM irrespective of the position of the crystal within the array, or the interaction location along the length of the crystal. The total light signal and energy resolution was almost independent of the depth of interaction. The measured energy resolution averaged 14% FWHM. The coincidence timing resolution measured using a pair of identical detector modules was 4.5 ns FWHM. These results are consistent with the design goals and the performance required of a compact, high-resolution and high-efficiency PET detector module for small animal and breast imaging applications
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/49/4293/pmb4_18_007.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/49/4293/pmb4_18_007.pdf; http://www.iop.org/;
- DOI
- 10.1088/0031-9155/49/18/007;
- PII
- S0031-9155(04)81065-0;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 49
- Journal Issue
- 18
- Journal Page Range
- p. 4293-4304
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 36032325
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AMPLITUDES; CRYSTALS; EFFICIENCY; ENERGY RESOLUTION; LUTETIUM; MAMMARY GLANDS; PERFORMANCE; PHOTODIODES; POSITRON COMPUTED TOMOGRAPHY; READOUT SYSTEMS; SCINTILLATIONS; SPATIAL RESOLUTION
- Descriptors DEC
- BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; ELEMENTS; EMISSION COMPUTED TOMOGRAPHY; GLANDS; METALS; ORGANS; RARE EARTHS; RESOLUTION; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; TOMOGRAPHY