A normalization technique for 3D PET data

  • 1. AZ-VUB, Brussels (Belgium). Div. of Nuclear Medicine
  • 2. University Hospital, Geneva (Switzerland). Div. of Nuclear Medicine
  • 3. Hammersmith Hospital, London (UK). M.R.C. Cyclotron Unit
  • 4. Positron Tomography Lab., Louvain-la-Neuve (Belgium)

Description

Prior to reconstruction emission data from a multi-ring PET camera must be corrected for detector sensitivity variations. The authors present an alternative to the standard calibration technique; by modelling the lines of response (LORs) sensitivity as the product of the individual detector efficiencies and a geometrical factor, a calibration procedure is proposed which requires the determination of only 6000 parameters from a data set. (author)