Published April 1990
| Version v1
Journal article
A new approach to very high resolution mini-brain PET using a small number of large detectors
- 1. California Univ., Irvine, CA (USA). Dept. of Radiological Sciences
- 2. Dept. of Psychiatry and Human Behavior, Univ. of California, Irvine, CA (USA)
- 3. Dept. of Ophthalmology, Univ. of California, Irvine, CA (USA)
Description
A new very high resolution positron emission tomography using only a small number of large detectors in a circular ring format with collimating apertures is proposed and a design example is given. The proposed system is ideally suited for very high resolution brain imaging, i.e., close to 2 mm fwhm resolution and can be converted to a medium resolution system by simply changing the size of collimating aperture if higher sensitivity operation is desired. The basic concept of the system design is the optimization of the detector size and system diameter. The former relates to the detection efficiency and photon spillage and penetration to adjacent crystals, while the latter (system diameter) relates to the overall system sensitivity
Additional details
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 37
- Journal Issue
- 2
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 842-851
- ISSN
- 0018-9499
- CODEN
- IETNA
Conference
- Title
- Institute for Electronic and Electrical Engineers (IEEE) nuclear science symposium.
- Dates
- 15-19 Jan 1990.
- Place
- San Francisco, CA (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 22004781
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- APERTURES; BRAIN; COLLIMATORS; DIAGNOSIS; FEASIBILITY STUDIES; IMAGE PROCESSING; IMAGE SCANNERS; POSITRON COMPUTED TOMOGRAPHY; RADIATION DETECTORS; SPECIFICATIONS
- Descriptors DEC
- BODY; CENTRAL NERVOUS SYSTEM; COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; MEASURING INSTRUMENTS; NERVOUS SYSTEM; OPENINGS; ORGANS; TOMOGRAPHY
Optional Information
- Secondary number(s)
- CONF-900143--.