Study on type-testing of a manual TLD-reader for dosimetry programs

  • 1. Nuclear Science and Technology Research Institute, Radiation Applications Research School, Tehran (Iran, Islamic Republic of)

Description

In a radiation individual monitoring program, the type testing of measuring devices is a great important part of the quality management system. The IEC-62387 standard applies to dosimetry systems that measure external photon and/or beta radiation within limited ranges of the associated physical parameters. In this work, a type-testing program was conducted for a manual thermoluminescence dosimetry reader employing the IEC-62387 radiation and environmental performance criteria. The uncertainty of non-linearity of the response of the dosimetry system in a range of 0.7-850 mSv was obtained between -15% and + 17%, which fulfilled the IEC standard range of -16% to +18%. Furthermore, the total uncertainty of all reader tests was measured to be 12%, which was less than the criteria of 20% in the IEC standard. Thus, it can be concluded that the thermoluminescence dosimetry reader met all requirements of the IEC standard for the reader-tests by an appropriate margin.

Availability note (English)

Available from Atomic Energy Organization of Iran

Additional details

Publishing Information

Journal Title
Journal of Nuclear Research and Applications
Journal Volume
2
Journal Issue
no.1
Journal Page Range
p. 61-67
ISSN
2783-3402

INIS

Country of Publication
Iran, Islamic Republic of
Country of Input or Organization
Iran, Islamic Republic of
INIS RN
55053694
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
CALIBRATION STANDARDS; DOSEMETERS; MEASURING INSTRUMENTS; PERFORMANCE TESTING; QUALITY CONTROL; RADIATION MONITORING; SSDL; THERMOLUMINESCENCE; THERMOLUMINESCENT DOSIMETRY
Descriptors DEC
CONTROL; DOSIMETRY; EMISSION; LUMINESCENCE; MEASURING INSTRUMENTS; MONITORING; PHOTON EMISSION; STANDARDS; TESTING