Published July 1999 | Version v1
Journal article

Investigation of the electronics nonlinearity in a HPGe γ spectrometer

  • 1. Academia Sinica, Beijing (China). Inst. of High Energy Physics
  • 2. China Inst. of Atomic Energy, Beijing (China)

Description

The nonlinearity deviation curve of each instrument module in a HPGe spectrometer has been tested by means of a precise pulse generator and then checked by the detector pulses induced by nuclear γ rays. A testing precision of 3 x 10-6 is achieved. The contribution from every component to the nonlinearity of whole spectrometer system is investigated

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
19
Journal Issue
4
Journal Page Range
p. 241-247
ISSN
0258-0934