Published September 1967 | Version v1
Book

Pre-Sterilization Bacterial Contamination on Disposable Hypodermic Syringes: Necessary Information for the Rational Choice of Dose for Radiation Sterilization

  • 1. Research Committee, UK Panel on Gamma and Electron Irradiation (United Kingdom)
  • 2. School of Pharmacy, University of London, London, W.C.I. (United Kingdom)
  • 3. Radiobiology Laboratory, Churchill Hospital, Oxford (United Kingdom)

Description

The numbers and types of micro-organisms contaminating disposable plastic hypodermic syringes have been assessed prior to sterilization. A large number of syringes have been taken at random from the production lines of three independent manufacturers. The microbiological investigations were carried out in the laboratories maintained by each of these manufacturers for routine sterility testing, according to a standard protocol devised by the Research Committee of the UK Panel on Gamma and Electron Irradiation, which co-ordinated the investigation and analysed the results. Items produced by any one manufacturer were assessed for microbial contamination both in their own laboratories and in the laboratories of the other manufacturers. The level of false positive' results was determined independently for each laboratory by the testing of known sterile' items which had been subjected to the radiation sterilization process. Both the percentage of syringes initially sterile, and the average number of organisms per contaminated syringe differed among the three manufacturers. When corrected for differences between the assessments by the various laboratories, the ranges observed were 16-48% initially sterile, and 20-70 organisms per contaminated syringe. Of 964 syringes tested by all three laboratories, only one contained over 1000 aerobic organisms (1133). The most common organisms found on the 62 more heavily contaminated syringes (100-1000 organisms) were coagulase-negative Gram-positive cocci. Two manufacturers assessed contamination by anaerobic organisms; on 610 syringes, one contained 4275 organisms, three more had 100- 1000 organisms, but 488 (80%) were uncontaminated by anaerobes. The results are discussed in the context of the choice of radiation dose necessary for the sterilization of medical products manufactured under controlled hygienic conditions. (author)

Part of:
Radiosterilization of Medical Products and Recommended Code of Practice. Proceedings of a Symposium

Additional details

Publishing Information

Publisher
IAEA
Imprint Place
Vienna (International Atomic Energy Agency (IAEA))
Imprint Title
Radiosterilization of Medical Products and Recommended Code of Practice. Proceedings of a Symposium
Imprint Pagination
481 p.
Series
Proceedings Series
Journal Page Range
p. 295-305
ISSN
0074-1884

Conference

Title
Symposium on Radiosterilization of Medical Products; Panel Meeting on a Code of Practice for the Radiosterilization of Medical Products
Dates
5-9 Jun 1967; 5-9 Dec 1966
Place
Budapest (Hungary); Vienna (Austria)

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46129912
Subject category
S60: APPLIED LIFE SCIENCES;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CONTAMINATION; MANUFACTURERS; PLASTICS; RADIATION DOSES; RADIOSTERILIZATION; STERILITY
Descriptors DEC
DOSES; IRRADIATION; MATERIALS; ORGANIC COMPOUNDS; ORGANIC POLYMERS; PETROCHEMICALS; PETROLEUM PRODUCTS; POLYMERS; STERILIZATION; SYNTHETIC MATERIALS

Optional Information

Notes
7 refs., 6 figs., 2 tabs.