Published March 1999
| Version v1
Journal article
Comparison of the NIST and BIPM air-kerma standards for measurements in the low-energy x-ray range
Creators
- 1. Bureau International des Poids et Mesures, Sevres (France)
- 2. National Inst. of Standards and Technology, Gaithersburg, MD (United States)
Description
A direct comparison was made between the air-kerma standards used for the measurement of low-energy x rays at the National Institute of Standards and Technology (NIST) and the Bureau International des Poids et Mesures (BIPM). The comparison was carried out at the BIPM using the BIPM reference beam qualities in the range from 10kV to 100 kV. The results show the standards to be in agreement to around 0.5% at reference beam qualities up to 50kV and at 100 kV. The result at the 80kV beam quality is less favorable, with agreement at the 1% level
Additional details
Publishing Information
- Journal Title
- Journal of Research of the National Institute of Standards and Technology
- Journal Volume
- 104
- Journal Issue
- 2
- Journal Page Range
- p. 135-140
- ISSN
- 1044-677X
- CODEN
- JRITEF
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 30051549
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- CALIBRATION STANDARDS; INTERLABORATORY COMPARISONS; IONIZATION CHAMBERS; KERMA; X-RAY DETECTION; X-RAY SOURCES
- Descriptors DEC
- DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RADIATION SOURCES; STANDARDS