Published March 1999 | Version v1
Journal article

Comparison of the NIST and BIPM air-kerma standards for measurements in the low-energy x-ray range

  • 1. Bureau International des Poids et Mesures, Sevres (France)
  • 2. National Inst. of Standards and Technology, Gaithersburg, MD (United States)

Description

A direct comparison was made between the air-kerma standards used for the measurement of low-energy x rays at the National Institute of Standards and Technology (NIST) and the Bureau International des Poids et Mesures (BIPM). The comparison was carried out at the BIPM using the BIPM reference beam qualities in the range from 10kV to 100 kV. The results show the standards to be in agreement to around 0.5% at reference beam qualities up to 50kV and at 100 kV. The result at the 80kV beam quality is less favorable, with agreement at the 1% level

Additional details

Publishing Information

Journal Title
Journal of Research of the National Institute of Standards and Technology
Journal Volume
104
Journal Issue
2
Journal Page Range
p. 135-140
ISSN
1044-677X
CODEN
JRITEF

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
30051549
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
CALIBRATION STANDARDS; INTERLABORATORY COMPARISONS; IONIZATION CHAMBERS; KERMA; X-RAY DETECTION; X-RAY SOURCES
Descriptors DEC
DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RADIATION SOURCES; STANDARDS