New CDF end-plug calorimeter

Description

The performance of the CDF end-plug calorimeter is described. The authors determined the energy resolutions to be σ/Ε = 14.5%/√Ε circleplus 0.7% for electron and σ/Ε = 68.0%/√Ε circleplus 4.1% for charged pion, where Ε is the energy measured in GeV. The response linearities satisfied their requirements. The response variations on the surface of a typical tower were measured to be 2.3% for e+ and 1.6% for π+. For the photon conversion detection by the preshower detector, they obtained the detection efficiency for two minimum ionizing particles to be 90--100% for the phototube gains of (1--4) x 105. The rate of π+'s depositing large fractions of energy in the electromagnetic section could be reduced by factors of 1.4--2.0 with keeping 95% efficiency for e+

Availability note (English)

Available from PURL: https://www.osti.gov/servlets/purl/751367-GPELFh/webviewable/

Additional details

Publishing Information

Imprint Pagination
7 p.
Report number
FERMILAB-Conf--00/036-E

Conference

Title
7. International Conference on Instrumentation for Colliding Beam Physics INSTR99
Dates
15-19 Nov 1999
Place
Hamamatsu (Japan)

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
32032235
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ELECTRON DETECTION; ENERGY RESOLUTION; FERMILAB COLLIDER DETECTOR; PERFORMANCE; PION DETECTION; SHOWER COUNTERS
Descriptors DEC
CHARGED PARTICLE DETECTION; DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION

Optional Information

Contract/Grant/Project number
AC02-76CH03000
Notes
This record replaces 31019642
Funding organization
USDOE Office of Energy Research (ER) (United States)