Description
The performance of the CDF end-plug calorimeter is described. The authors determined the energy resolutions to be σ/Ε = 14.5%/√Ε circleplus 0.7% for electron and σ/Ε = 68.0%/√Ε circleplus 4.1% for charged pion, where Ε is the energy measured in GeV. The response linearities satisfied their requirements. The response variations on the surface of a typical tower were measured to be 2.3% for e+ and 1.6% for π+. For the photon conversion detection by the preshower detector, they obtained the detection efficiency for two minimum ionizing particles to be 90--100% for the phototube gains of (1--4) x 105. The rate of π+'s depositing large fractions of energy in the electromagnetic section could be reduced by factors of 1.4--2.0 with keeping 95% efficiency for e+
Availability note (English)
Available from PURL: https://www.osti.gov/servlets/purl/751367-GPELFh/webviewable/Additional details
Identifiers
Publishing Information
- Imprint Pagination
- 7 p.
- Report number
- FERMILAB-Conf--00/036-E
Conference
- Title
- 7. International Conference on Instrumentation for Colliding Beam Physics INSTR99
- Dates
- 15-19 Nov 1999
- Place
- Hamamatsu (Japan)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 32032235
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- ELECTRON DETECTION; ENERGY RESOLUTION; FERMILAB COLLIDER DETECTOR; PERFORMANCE; PION DETECTION; SHOWER COUNTERS
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION
Optional Information
- Contract/Grant/Project number
- AC02-76CH03000
- Notes
- This record replaces 31019642
- Funding organization
- USDOE Office of Energy Research (ER) (United States)