Published 1997
| Version v1
Journal article
Spectrometry of the linear energy transfer in high energy reference fields
- 1. Department of Radiation Dosimetry, Institute of Nuclear Physics, Academy of Sciences of the Czech Republic, Prague (Czech Republic)
- 2. Department of Radiation and Radiobiological Research, Joint Institute of Nuclear Research, Dubna (Russian Federation)
Description
Measurements of the high energy reference fields established behind the shielding of accelerators at CERN, Geneva, and JINR, Dubna, were made using a Linear Energy Transfer (LET) spectrometer based on a chemically etched track detector. The experiments demonstrated the wide capabilities of this microdosimetric technique. The LET spectra are in a reasonable agreement with those obtained by means of a tissue equivalent proportional counter. The integral values of the dosimetric characteristics calculated from the LET spectra also approach closely those obtained by other dosimetric and microdosimetric methods. (P.A.)
Additional details
Publishing Information
- Journal Title
- Bezpecnost Jaderne Energie
- Journal Volume
- 5
- Journal Issue
- 11-12
- Journal Page Range
- p. 354-358.
- ISSN
- 1210-7085
- CODEN
- BJENE3
INIS
- Country of Publication
- Czech Republic
- Country of Input or Organization
- Czech Republic
- INIS RN
- 29013988
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CALIBRATION; DIELECTRIC TRACK DETECTORS; LET; MICRODOSIMETRY; NEUTRON SPECTRA; NEUTRON SPECTROSCOPY; PROTONS; RADIATION DOSE DISTRIBUTIONS; SPECTROMETERS
- Descriptors DEC
- BARYONS; CATIONS; CHARGED PARTICLES; DOSIMETRY; ELEMENTARY PARTICLES; ENERGY TRANSFER; FERMIONS; HADRONS; HYDROGEN IONS; HYDROGEN IONS 1 PLUS; IONS; MEASURING INSTRUMENTS; NUCLEONS; RADIATION DETECTORS; SPECTRA; SPECTROSCOPY
Optional Information
- Notes
- 3 tabs., 3 figs., 14 refs.