Published October 1978 | Version v1
Journal article

Volumetric calibration of experimental apparatus by xenon gas transfer

  • 1. Monsanto Research Corporation, Mound Laboratory, Miamisburg, Ohio 45342

Description

The volumes of interconnected sections of large, irregular, or inaccessible apparatus can be measured by quantitative cryogenic transfer of a xenon gas fill to weighing bottles. Tests with eight containers from 75 to 10 000 cm3 using relatively simple equipment, showed that: (1) absolute volumes of 250 cm3 or larger can be measured with a reproducibility on the order of 0.02% and with an accuracy on the order of 0.05% and (2) the ratios among two or more volumes can be determined with a reproducibility of 0.01%--0.001%, depending on volume

Additional details

Publishing Information

Journal Title
Rev. Sci. Instrum.
Journal Volume
49
Journal Issue
10
Series
Rev. Sci. Instrum.
Journal Page Range
1415-1417

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
10423684
Subject category
S42: ENGINEERING;
Descriptors DEI
ACCURACY; CALIBRATION; ERRORS; GAS FLOW; MEASURING METHODS; VOLUME; XENON
Descriptors DEC
ELEMENTS; FLUID FLOW; NONMETALS; RARE GASES