Published October 1978
| Version v1
Journal article
Volumetric calibration of experimental apparatus by xenon gas transfer
Creators
- 1. Monsanto Research Corporation, Mound Laboratory, Miamisburg, Ohio 45342
Description
The volumes of interconnected sections of large, irregular, or inaccessible apparatus can be measured by quantitative cryogenic transfer of a xenon gas fill to weighing bottles. Tests with eight containers from 75 to 10 000 cm3 using relatively simple equipment, showed that: (1) absolute volumes of 250 cm3 or larger can be measured with a reproducibility on the order of 0.02% and with an accuracy on the order of 0.05% and (2) the ratios among two or more volumes can be determined with a reproducibility of 0.01%--0.001%, depending on volume
Additional details
Publishing Information
- Journal Title
- Rev. Sci. Instrum.
- Journal Volume
- 49
- Journal Issue
- 10
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 1415-1417
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 10423684
- Subject category
- S42: ENGINEERING;
- Descriptors DEI
- ACCURACY; CALIBRATION; ERRORS; GAS FLOW; MEASURING METHODS; VOLUME; XENON
- Descriptors DEC
- ELEMENTS; FLUID FLOW; NONMETALS; RARE GASES