Mass characterization of multi-pixel photon counters for the T2K 280 m near detector

  • 1. Laboratoire Leprince-Ringuet-Ecole polytechnique, CNRS/IN2P3 (LLR), 91128 Palaiseau, CNRS (France)

Description

A few thousands multi-pixel photon counters (MPPC) have been characterized for the T2K 280 m near detector (ND280). For this purpose a test facility has been developed enabling the characterization of 128 MPPC per day. A mean gain value of 5.4x105 at breakdown voltage (Vbreak)+1 V with a dispersion of 2.6% has been measured at 20 deg. C. It is shown that the accuracy of the test facility for the gain measurement is at the level of 0.76%. At 20 deg. C the dark noise rates (DNR) have been measured in a range from ∼400 up to ∼730 kHz at Vbreak+1.6 V.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2009.11.042

Additional details

Identifiers

DOI
10.1016/j.nima.2009.11.042;
PII
S0168-9002(09)02222-0;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
613
Journal Issue
1
Journal Page Range
p. 46-53
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41083133
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ACCURACY; BREAKDOWN; ELECTRIC POTENTIAL; GAIN; KHZ RANGE 100-1000; MASS; NOISE; PHOTONS; QUALITY CONTROL; TEST FACILITIES
Descriptors DEC
AMPLIFICATION; BOSONS; CONTROL; ELEMENTARY PARTICLES; FREQUENCY RANGE; KHZ RANGE; MASSLESS PARTICLES