Dose verification by OSLDs in the irradiation of cell cultures
Creators
- 1. Ce. De. Te. Centro Especializado en Terapia Radiante, Viamonte 1861, C1056 Buenos Aires (Argentina)
- 2. Comision Nacional de Energia Atomica, Departamento de Micro y Nanotecnologia, Av. Gral. Paz 1499, 1650 San Martin, Buenos Aires (Argentina)
Description
The determination of value of irradiation dose presents difficulties when targets are irradiated located in regions where electronic equilibrium of charged particle is not reached, as in the case of irradiation -in vitro- of cell lines monolayer-cultured, in culture dishes or flasks covered with culture medium. The present study aimed to implement a methodology for dose verification in irradiation of cells in culture media by optically stimulated luminescence dosimetry (OSLD). For the determination of the absorbed dose in terms of cell proliferation OSL dosimeters of aluminum oxide doped with carbon (Al2O3:C) were used, which were calibrated to the irradiation conditions of culture medium and at doses that ranged from 0.1 to 15 Gy obtained with a linear accelerator of 6 MV photons. Intercomparison measurements were performed with an ionization chamber of 6 cm3. Different geometries were evaluated by varying the thicknesses of solid water, air and cell culture medium. The results showed deviations below 2.2% when compared with the obtained doses of OSLDs and planning system used. Also deviations were observed below 3.4% by eccentric points of the irradiation plane, finding homogeneous dose distribution. Uncertainty in the readings was less than 2%. The proposed methodology contributes a contribution in the dose verification in this type of irradiations, eliminating from the calculation uncertainties, potential errors in settling irradiation or possible equipment failure with which is radiating. It also provides certainty about the survival curves to be plotted with the experimental data. (Author)
Files
47032311.pdf
Files
(426.9 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:89db03fab123e7d95180e4aa0ce66c83
|
426.9 kB | Preview Download |
Additional details
Additional titles
- Original title (Spanish)
- Verificacion de dosis mediante OSLDs, en la irradiacion de cultivos celulares
Publishing Information
- Publisher
- Sociedad Mexicana de Irradiacion y Dosimetria
- Imprint Place
- Mexico, D. F. (Mexico)
- Imprint Pagination
- 8 p.
- Report number
- INIS-MX--2975
Conference
- Title
- 15. International Symposium on Solid State Dosimetry
- Original Conference Title
- 15. Conferencia Internacional sobre Dosimetria de Estado Solido
- Dates
- 26-30 Sep 2015
- Place
- Leon, Guanajuato (Mexico)
INIS
- Country of Publication
- Mexico
- Country of Input or Organization
- Mexico
- INIS RN
- 47032311
- Subject category
- S63: RADIATION, THERMAL, AND OTHER ENVIRONMENTAL POLLUTANT EFFECTS ON LIVING ORGANISMS AND BIOLOGICAL MATERIALS;
- Resource subtype / Literary indicator
- Conference, Numerical Data
- Descriptors DEI
- ABSORBED RADIATION DOSES; AIR; ALUMINIUM OXIDES; CARBON; CELL CULTURES; CELL PROLIFERATION; COMPARATIVE EVALUATIONS; CULTURE MEDIA; DOPED MATERIALS; DOSEMETERS; EXPERIMENTAL DATA; GEOMETRY; IRRADIATION; LINEAR ACCELERATORS; LUMINESCENCE; PHOTONS; SURVIVAL CURVES; THICKNESS; VERIFICATION; WATER
- Descriptors DEC
- ACCELERATORS; ALUMINIUM COMPOUNDS; BOSONS; CHALCOGENIDES; DATA; DIMENSIONS; DOSES; ELEMENTARY PARTICLES; ELEMENTS; EMISSION; EVALUATION; FLUIDS; GASES; HYDROGEN COMPOUNDS; INFORMATION; MASSLESS PARTICLES; MATERIALS; MATHEMATICS; MEASURING INSTRUMENTS; NONMETALS; NUMERICAL DATA; OXIDES; OXYGEN COMPOUNDS; PHOTON EMISSION; RADIATION DOSES
Optional Information
- Funding organization
- National Council for Science and Technology (Mexico); International Centre for Theoretical Physics (Italy); Asesores en Proteccion Radiologica y Nuclear, S. C. (Mexico); DOSImetrics (Germany); RadMedical (Mexico); Convention and Visitors Bureau of Leon (Mexico); Tecnofisica, S. A. de C. V. (Mexico)