Published February 1983 | Version v1
Report Restricted

Calculation of the performance of the INS iron-free π√2 spectrometer as a spectrograph

  • 1. Tohoku Univ., Sendai (Japan). Dept. of Physics

Description

The performance of the INS iron-free π√2 beta-ray spectrometer of the current-loop type is calculated with a view of using it as a spectrograph, i.e., in a multichannel mode with a position-sensitive proportional counter. For the momentum resolution of R = 0.01 and 0.1 % the usable momentum range as a spectrograph (+epsilon+ < 2.5 and 5%, respectively) and the optimum baffle position (phi sub(B) = 250) and the line shapes on the focal plane are calculated. The transmission of the baffle is 0.025 and 0.13 % of 4π and the expected gain of data-collection efficiency over the single-channel mode is 140 and 40 for R = 0.01 and 0.1%, respectively. An effective tilting of the focal plane due to the entrance baffle is discussed as well as the problems with arrangement and testing of the position detector. (author)

Availability note (English)

MF available from INIS under the Report Number.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
23 p.
Report number
INS-J--166