A preliminary study on the determination of boron and cadmium impurities in uranium oxide and uranium metals by Inductively Coupled Plasma-Atomic Emission Spectroscopy

  • 1. Ames Lab. - USDOE, Iowa State Univ., Ames, IA 50011

Description

Boron and cadmium impurities in U metals and U3O8 may be determined by inductively coupled plasma-atomic emission spectroscopy (ICP-AES) after their separation from the U matrix by multiple extractions with tri-n-octyl phosphine oxide (TOPO). Useful precision and accuracy were obtained at concentrations of approx. 0.2 μg/g U

Additional details

Publishing Information

Publisher
Elsevier.
Imprint Place
New York, NY (USA)
Imprint Title
Analytical spectroscopy
Journal Page Range
p. 187-193.

Conference

Title
26. ORNL-DOE conference on analytical chemistry in energy technology.
Dates
11-13 Oct 1983.
Place
Knoxville, TN (USA).