Uncertainty analysis in the simulation of an HPGe detector using the Monte Carlo Code MCNP5
Creators
- 1. Universitat Politecnica de Valencia, Valencia, (Spain). Instituto de Seguridad Industrial, Radiofisica y Medioambiental (ISIRYM)
- 2. Universitat Politecnica de Valencia, Valencia, (Spain). Servicio de Radiaciones. Lab. de Radiactividad Ambiental
- 3. Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Engenharia Nuclear
Description
A gamma spectrometer including an HPGe detector is commonly used for environmental radioactivity measurements. Many works have been focused on the simulation of the HPGe detector using Monte Carlo codes such as MCNP5. However, the simulation of this kind of detectors presents important difficulties due to the lack of information from manufacturers and due to loss of intrinsic properties in aging detectors. Some parameters such as the active volume or the Ge dead layer thickness are many times unknown and are estimated during simulations. In this work, a detailed model of an HPGe detector and a petri dish containing a certified gamma source has been done. The certified gamma source contains nuclides to cover the energy range between 50 and 1800 keV. As a result of the simulation, the Pulse Height Distribution (PHD) is obtained and the efficiency curve can be calculated from net peak areas and taking into account the certified activity of the source. In order to avoid errors due to the net area calculation, the simulated PHD is treated using the GammaVision software. On the other hand, it is proposed to use the Noether-Wilks formula to do an uncertainty analysis of model with the main goal of determining the efficiency curve of this detector and its associated uncertainty. The uncertainty analysis has been focused on dead layer thickness at different positions of the crystal. Results confirm the important role of the dead layer thickness in the low energy range of the efficiency curve. In the high energy range (from 300 to 1800 keV) the main contribution to the absolute uncertainty is due to variations in the active volume. (author)
Files
45107398.pdf
Files
(210.0 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:839863c29f5a1d753aae15453166f4ec
|
210.0 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 12 p.
- Report number
- INIS-BR--14568
Conference
- Title
- international nuclear atlantic conference
- Acronym
- INAC 2013
- Dates
- 24-29 Nov 2013
- Place
- Recife, PE (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 45107398
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPUTERIZED SIMULATION; DATA COVARIANCES; GAMMA SOURCES; GAMMA SPECTROMETERS; HIGH-PURITY GE DETECTORS; LAYERS; MONTE CARLO METHOD; POLYNOMIALS; RADIATION DETECTION; THICKNESS
- Descriptors DEC
- CALCULATION METHODS; DETECTION; DIMENSIONS; FUNCTIONS; GE SEMICONDUCTOR DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATION SOURCES; SEMICONDUCTOR DETECTORS; SIMULATION; SPECTROMETERS
Optional Information
- Notes
- 1 ref., 7 figs., 2 tabs.