Consideration is given to X-ray fluorescence analyzer for analyzing pulps and solutions immediately in the flow and to results of its performance testing. The analyzer contains two dip sensors, the unit for analogue data processing and recording unit. The relative root-mean-square error of zirconium determination in 0.29-4.77 g/l range was equal to about 10% for 100 s measuring period. Zirconium distribution with the depth of analyzed volume and the effect of pulp particle size on analysis results were investigated

Files

19095624.pdf

Files (701.8 kB)

Name Size Download all
md5:5ef32df9d37e46af6f8a7c87a4df855a
701.8 kB Preview Download

System files (20.5 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Рентгенорадиометрический анализ циркониевых пульп погружными датчиками

Publishing Information

Imprint Title
Radiation technology
Journal Issue
no.2(35
Series
Voprosy atomnoj nauki i tekhniki.
Journal Page Range
p. 20-22.
Report number
INIS-SU--58

Optional Information

Notes
3 refs.; 4 figs. Imprint:Radiatsionnaya tekhnika.;Nauchno-tekhnicheskij sbornik.