SU-D-206-06: Task-Specific Optimization of Scintillator Thickness for CMOS-Detector Based Cone-Beam Breast CT
- 1. University of Massachusetts Medical School, Worcester, MA (United States)
Description
Purpose: To optimize the cesium iodide (CsI:Tl) scintillator thickness in a complimentary metal-oxide semiconductor (CMOS)-based detector for use in dedicated cone-beam breast CT. Methods: The imaging task considered was the detection of a microcalcification cluster comprising six 220µm diameter calcium carbonate spheres, arranged in the form of a regular pentagon with 2 mm spacing on its sides and a central calcification, similar to that in ACR-recommended mammography accreditation phantom, at a mean glandular dose of 4.5 mGy. Generalized parallel-cascades based linear systems analysis was used to determine Fourier-domain image quality metrics in reconstructed object space, from which the detectability index inclusive of anatomical noise was determined for a non-prewhitening numerical observer. For 300 projections over 2π, magnification-associated focal-spot blur, Monte Carlo derived x-ray scatter, K-fluorescent emission and reabsorption within CsI:Tl, CsI:Tl quantum efficiency and optical blur, fiberoptic plate transmission efficiency and blur, CMOS quantum efficiency, pixel aperture function and additive noise, and filtered back-projection to isotropic 105µm voxel pitch with bilinear interpolation were modeled. Imaging geometry of a clinical prototype breast CT system, a 60 kV Cu/Al filtered x-ray spectrum from 0.3 mm focal spot incident on a 14 cm diameter semi-ellipsoidal breast were used to determine the detectability index for 300–600 µm thick (75µm increments) CsI:Tl. The CsI:Tl thickness that maximized the detectability index was considered optimal. Results: The limiting resolution (10% modulation transfer function, MTF) progressively decreased with increasing CsI:Tl thickness. The zero-frequency detective quantum efficiency, DQE(0), in projection space increased with increasing CsI:Tl thickness. The maximum detectability index was achieved with 525µm thick CsI:Tl scintillator. Reduced MTF at mid-to-high frequencies for 600µm thick CsI:Tl lowered the detectability index than 525µm CsI:Tl. Conclusion: For the x-ray spectrum and imaging conditions considered, a 525µm thick CsI:Tl scintillator integrated with the CMOS detector is optimal for detecting microcalcification cluster. Funding support: Supported in part by NIH R01 CA195512. The contents are solely the responsibility of the authors and do not reflect the official views of the NIH or the NCI. Disclosures: SV, GV and AK - Research collaboration, Koning Corp., West Henrietta, NY.
Additional details
Identifiers
- DOI
- 10.1118/1.4955660;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 43
- Journal Issue
- 6
- Journal Page Range
- vp.
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48071898
- Subject category
- S60: APPLIED LIFE SCIENCES; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CESIUM IODIDES; COMPUTERIZED TOMOGRAPHY; FLUORESCENCE; INTERPOLATION; MAMMARY GLANDS; MONTE CARLO METHOD; PHANTOMS; QUANTUM EFFICIENCY; SEMICONDUCTOR MATERIALS; THICKNESS; TRANSFER FUNCTIONS; X-RAY SPECTRA
- Descriptors DEC
- ALKALI METAL COMPOUNDS; BODY; CALCULATION METHODS; CESIUM COMPOUNDS; CESIUM HALIDES; DIAGNOSTIC TECHNIQUES; DIMENSIONS; EFFICIENCY; EMISSION; FUNCTIONS; GLANDS; HALIDES; HALOGEN COMPOUNDS; INORGANIC PHOSPHORS; IODIDES; IODINE COMPOUNDS; LUMINESCENCE; MATERIALS; MATHEMATICAL SOLUTIONS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; NUMERICAL SOLUTION; ORGANS; PHOSPHORS; PHOTON EMISSION; RADIOLOGY; SPECTRA; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- (c) 2016 American Association of Physicists in Medicine