DXRaySMCS: A user-friendly interface developed for prediction of diagnostic radiology X-ray spectra produced by Monte Carlo (MCNP-4C) simulation
- 1. Medical Physics Research Center, Mashhad Univ. of Medical Sciences, Mashhad (Iran, Islamic Republic of)
- 2. Medical Physics and Technology of Radiology Dept., Faculty of Para Medical Sciences, Kashan Univ. of Medical Sciences, Kashan (Iran, Islamic Republic of)
- 3. Dept. of Radiological Technology, Faculty of Para Medical Sciences, Mashhad (Iran, Islamic Republic of)
Description
In this work, the general purpose Monte Carlo N-particle radiation transport computer code (MCNP-4C) was used for the simulation of X-ray spectra in diagnostic radiology. The electron's path in the target was followed until its energy was reduced to 10 keV. A user-friendly interface named 'diagnostic X-ray spectra by Monte Carlo simulation (DXRaySMCS)' was developed to facilitate the application of MCNP-4C code for diagnostic radiology spectrum prediction. The program provides a user-friendly interface for: (i) modifying the MCNP input file, (ii) launching the MCNP program to simulate electron and photon transport and (iii) processing the MCNP output file to yield a summary of the results (relative photon number per energy bin). In this article, the development and characteristics of DXRaySMCS are outlined. As part of the validation process, output spectra for 46 diagnostic radiology system settings produced by DXRaySMCS were compared with the corresponding IPEM78. Generally, there is a good agreement between the two sets of spectra. No statistically significant differences have been observed between IPEM78 reported spectra and the simulated spectra generated in this study. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncn312Additional details
Identifiers
- DOI
- 10.1093/rpd/ncn312;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 132
- Journal Issue
- 4
- Journal Page Range
- p. 415-419
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43120720
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S97: MATHEMATICAL METHODS AND COMPUTING; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTERIZED SIMULATION; FORECASTING; MONTE CARLO METHOD; PHOTON TRANSPORT; RADIOLOGY; X-RAY SPECTRA
- Descriptors DEC
- CALCULATION METHODS; MEDICINE; NEUTRAL-PARTICLE TRANSPORT; NUCLEAR MEDICINE; RADIATION TRANSPORT; SIMULATION; SPECTRA
Optional Information
- Notes
- 11 refs