Published November 2000 | Version v1
Journal article

New duct caliber measurement methods on magnetic resonance cholangiopancreatography

  • 1. Tohoku Univ., Sendai, Miyagi (Japan). School of Medicine

Description

The purpose of this study is to report the feasibility of new duct caliber measurement methods using signal intensity on MR cholangiopancreatography (MRCP). A phantom with 0.25-8 mm caliber ducts was imaged with a 1.5-Tesla MR system using the usual MRCP sequence. Each caliber was measured by conventional method (full width at half maximum, FWHM) and two new methods: the maximum intensity measurement method (MIM), in which caliber was calculated from maximum signal intensity, and the area intensity measurement method (AIM), in which caliber was calculated from the area under the signal intensity curve. Errors in calculated caliber were compared among the three methods of measurement. Mean measurement errors of each caliber for 1-8 mm ducts by MIM [14% (2.1-25%)] and by AIM [6.8% (0.4-16%)] were significantly lower than those by FWHM [30% (17-64%)]. AIM was significantly more accurate than MIM for caliber measurement of ducts over 1 mm. Two new caliber measurement methods for pancreatobiliary ducts in MRCP, MIM and AIM are accurate and should play an important role in clinical MRCP. (author)

Additional details

Publishing Information

Journal Title
Nippon Igaku Hoshasen Gakkai Zasshi
Journal Volume
60
Journal Issue
13
Journal Page Range
p. 738-745
ISSN
0048-0428