Effects of anisotropic fluences and angular depended spectra of beta-particles in the use of large area reference sources
Creators
- 1. Japan Radioisotope Association, 2-28-45, Hon-komagome, Bunkyo, Tokyo 113-8941 (Japan)
- 2. National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Tsukuba Central 2, 1-1-1 Umezono, Ttsukuba, Ibaraki 305-8568 (Japan)
Description
Calibrations of instrument efficiency of surface contamination meters are usually made with extended reference sources which are standardized in terms of 2π surface β-particle emission rates from the source surface including backscattered particles. Extended sources supplied from various metrology institutes or calibration laboratories, but the source-types such as structure, preparation method, backing and covering materials vary between manufacturers. In this work first we show how the calibration results are dependent on the source type. Second, in order to clarify the possible reason of such discrepancy, we examined the isotropy of β-particle fluences by the use of a proportional counter and also observed the angular dependence of β-particle spectra by the use of small plastic scintillation spectrometer, where the source mount can rotate relative to the detector window at various obliquities. The discrepancy in the instrument-calibration of surface contamination meters, which are mainly used under the conditions of large source-to-detector geometry, can be explained. - Highlights: ► We show how the calibration results are dependent on the source type. ► We examined the isotropy of β-particle fluences and observed the angular dependence. ► Discrepancy of instrument efficiencies using different type of sources is explained.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.apradiso.2012.02.037Additional details
Identifiers
- DOI
- 10.1016/j.apradiso.2012.02.037;
- PII
- S0969-8043(12)00085-1;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 70
- Journal Issue
- 9
- Journal Page Range
- p. 1959-1963
- ISSN
- 0969-8043
- CODEN
- ARISEF
Conference
- Title
- 18. international conference on radionuclide metrology and its applications
- Dates
- 19-23 Sep 2011
- Place
- Tsukuba (Japan)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44101163
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ANISOTROPY; BETA PARTICLES; CALIBRATION; EFFICIENCY; EMISSION; ISOTROPY; PLASTIC SCINTILLATION DETECTORS; PROPORTIONAL COUNTERS; SPECTRA; SPECTROMETERS; SURFACE CONTAMINATION
- Descriptors DEC
- CHARGED PARTICLES; CONTAMINATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS
Optional Information
- Copyright
- Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.