Published April 2010
| Version v1
Journal article
Fast integrator based data acquisition system for the SST-1 Thomson scattering system
Creators
- 1. Institute for Plasma Research, Bhat, Gandhinagar 382 428 (India)
Description
An operational transconductance amplifier based fast charge-integrating module (FCIM) is designed and developed for an easy acquisition of fast Thomson scattered and background signal. FCIM based data acquisition technique can be used for the measurement of charge pulses of <20 ns duration. The response of the module is tested using a standard pulsed charge-generating module. The measuring charge range of the module depends on the integrating capacitor. Comparison of the performance of FCIM to commercially available module shows that it has better dynamic range with higher sensitivity and less measurement error. The module is quite cost effective and has many new features.
Additional details
Identifiers
- DOI
- 10.1063/1.3374110;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 81
- Journal Issue
- 4
- Journal Page Range
- p. 043501-043501.5
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44013940
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- CAPACITORS; DATA ACQUISITION SYSTEMS; OPERATIONAL AMPLIFIERS; SENSITIVITY; THOMSON SCATTERING; TOKAMAK DEVICES
- Descriptors DEC
- AMPLIFIERS; CLOSED PLASMA DEVICES; ELECTRICAL EQUIPMENT; ELECTRONIC EQUIPMENT; EQUIPMENT; INELASTIC SCATTERING; SCATTERING; THERMONUCLEAR DEVICES
Optional Information
- Notes
- (c) 2010 American Institute of Physics