Published May 1978
| Version v1
Report
Restricted
Gross gamma-ray calibration blocks. Part I. Gross gamma-ray calibration blocks; construction. Part II. Gross gamma-ray calibration blocks; grade assignment
Description
Eleven transportable, portable hand counter or face scanner calibration blocks were constructed from predetermined mixtures of sand, uranium ore, and concrete. These blocks, 42 in. dia. and 17 in. deep, approximate an infinite source, in both the vertical and horizontal directions, for the gamma rays of interest. The grades assigned to these blocks, for the purpose of calibrating portable gross gamma detectors, were determined by a procedure based on a standard least-squares technique for limiting sampling errors
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS., PC A05/MF A01.
Files
Additional details
Publishing Information
- Imprint Pagination
- 97 p.
- Report number
- GJBX--59(78)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 10429931
- Subject category
- S58: GEOSCIENCES; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION STANDARDS; GAMMA DETECTION; RADIATION DETECTORS; URANIUM ORES
- Descriptors DEC
- MEASURING INSTRUMENTS; ORES; RADIATION DETECTION