Description

99mTc-DTPA preparations were analyzed using an abbreviated two-column chromatographic system. Separate columns of Sephadex/saline and alumina/acetone were utilized in sequence to separate 99mTc-DTPA from both 99mTcO4- and "hydrolyzed-99mTc." Control studies showed that small Sephadex columns do remove "hydrolyzed-99mTc" from a test sample, alumina is capable of binding 99mTc-DTPA, and neither column has an affinity for 99mTcO4-. The mini column approach has the potential for use as a routine quality control test method.