Quality assurance of electron beams using a Varian electronic portal imaging device
Creators
- 1. Department of Radiation Physics, Princess Margaret Cancer Centre, University Health Network, Toronto, Ontario (Canada)
Description
The feasibility of utilizing an electronic portal imaging device (EPID) for the quality assurance of electron beams was investigated. This work was conducted on a Varian 2100iX machine equipped with an amorphous silicon (aS1000) portal imager. The linearity of the imager pixel response as a function of exposed dose was first confirmed. The short-term reproducibility of the EPID response to electron beams was verified. Low (6 MeV), medium (12 MeV) and high (20 MeV) energies were tested, each along with small (6 × 6 cm2), medium (10 × 10 cm2) and large (20 × 20 cm2) applicators. Acquired EPID images were analyzed using an in-house MATLAB code for radiation field size, penumbra, symmetry and flatness. Field sizes and penumbra values agreed with those from film dosimetry to within 1 mm. Field symmetry and flatness constancies were measured over a period of three weeks. The results indicate that EPID can be used for routine quality assurance of electron beams. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/58/16/5461Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 58
- Journal Issue
- 16
- Journal Page Range
- p. 5461-5475
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44076498
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ELECTRON BEAMS; FILM DOSIMETRY; IMAGE PROCESSING; IMAGES; QUALITY ASSURANCE; SILICON
- Descriptors DEC
- BEAMS; DOSIMETRY; ELEMENTS; LEPTON BEAMS; PARTICLE BEAMS; PROCESSING; SEMIMETALS